This page (revision-14) was last changed on 27-Nov-2017 16:38 by Peter Young

This page was created on 08-Oct-2013 20:24 by Peter Young

Only authorized users are allowed to rename pages.

Only authorized users are allowed to delete pages.

Page revision history

Version Date Modified Size Author Changes ... Change note
14 27-Nov-2017 16:38 4 KB Peter Young to previous
13 07-Dec-2016 14:14 3 KB ogenopanurequ to previous | to last
12 03-Nov-2016 18:27 5 KB eirjiyu to previous | to last
11 03-Nov-2016 18:26 4 KB ajeekinok to previous | to last
10 01-Nov-2016 16:44 3 KB aipofazakefi to previous | to last
9 29-Oct-2013 21:52 3 KB Peter Young to previous | to last
8 23-Oct-2013 11:33 3 KB Harry Warren to previous | to last
7 21-Oct-2013 12:54 3 KB Harry Warren to previous | to last
6 21-Oct-2013 12:50 2 KB Harry Warren to previous | to last
5 21-Oct-2013 12:40 2 KB Harry Warren to previous | to last
4 21-Oct-2013 12:33 1 KB Harry Warren to previous | to last
3 21-Oct-2013 12:31 1 KB Harry Warren to previous | to last
2 09-Oct-2013 14:47 1 KB Peter Young to previous | to last
1 08-Oct-2013 20:24 1 KB Peter Young to last

Page References

Incoming links Outgoing links

Version management

Difference between version and

At line 50 added 24 lines
!!Applying the new calibrations to the output from EIS_AUTO_FIT
If you use the Gaussian fitting routine EIS_AUTO_FIT, then the recommendation is to apply the new calibrations at the EIS_GET_FITDATA stage. For example,
{{IDL> int=eis_get_fitdata(fitdata,/int,calib=1}}
will apply the Del Zanna (2013) calibration to the output intensity. The options are:
0 - the calibration is not changed\\
1 - the Del Zanna (2013) calibration is applied\\
2 - /correct_sensitivity is "undone" and the Del Zanna (2013) calibration is applied\\
3 - the Warren et al. (2014) calibration is applied\\
4 - /correct_sensitivity is "undone" and the Warren et al. (2014) calibration is applied
To summarize, the steps to be done are:
* Use EIS_PREP to calibrate the level-0 file with standard options (no /correct_sensitivity).
* Extract the windata structure using EIS_GETWINDATA with the /refill option.
* Perform the Gaussian fit with EIS_AUTO_FIT.
* Extract the intensity array with EIS_GET_FITDATA using the calib option set to 1 or 3 (Del Zanna or Warren).